Guidance and manufacturer’s declaration-electromagnetic immunity-
for all EQUIPMENT AND SYSTEMS
Guidance and manufacturer’s declaration-electromagnetic immunity
The ET2400LM is intended for use in the electromagnetic environment specified below. The customer or the user
of the ET2400LM should assure that it is used in such an environment.
Immunity test IEC 60601 Compliance level Electromagnetic environment-
test level guidelines
Electrostatic ± 6 kV contact ± 6 kV contact Floors should be wood, concrete or
discharge(ESD) ± 8 kV air ± 8 kV air ceramic tile. If floors are covered with
IEC 61000-4-2 synthetic material, the relative humidity
should be at least 30%.
Electrical Fast ± 2 kV for power ± 2 kV for power Mains power quality should be that of a
transient/burst supply lines supply lines typical commerical or hospital
environment.
IEC 61000-4-4 ± 1 kV for input/output ± 1 kV for input/output
lines lines
Surge ± 1 kV line(s) to line(s) ± 1 kV line(s) to line(s) Mains power quality should be that of a
IEC 61000-4-5 ± 2 kV line(s) to earth ± 2 kV line(s) to earth typical commerical or hospital
environment.
Voltage dips, short <5% U
T <5% UT Mains power quality be that of a
interruption and (>95% dip in UT) (>95% dip in UT) typical commerical or hospital
voltage variations for 0.5 cycle for 0.5 cycle environment. If the user of the ET2400LM
on power supply requires continued operation during power
input lines 40% U
T 40% UT mains interruptions, it is recommended that
(60% dip in UT) (60 % dip in UT) the ET2400LM be powered from an
IEC 61000-4-11 for 5 cycles for 5 cycles uninterruptible power supply or a battery.
70% U
T 70% UT
(30% dip in UT) (30% dip in UT)
for 25 cycles for 25 cycles
<5% U
T <5% UT
(>95% dip in UT) (>95% dip in UT)
for 25 cycles for 25 cycles
Power frequency 3 A/m 3A/m Power frequency magnetic fields
(50/60 Hz) should be at levels characteristic of a
magnetic field typical location in a typical commerical
or hospital environment.
IEC 61000-4-8
NOTE U
T is the a.c. mains voltage prior to application of the test level.