Guidance and manufacturer’s declaration-electromagnetic immunity
for all EQUIPMENT AND
SYSTEMS
Guidance and manufacturer’s declaration-electromagnetic immunity
The 2401LM Touchmonitor for Healthcare Applications is intended for use in the electromagnetic
environment specified below. The customer, or the user of the 2401LM Touchmonitor for Healthcare
Applications,
should assure that it is used in such an environment.
Immunity Test
Level
IEC
60601
Compl
iance Level
Electromagnetic Environment - Guidelines
Electrostatic
Discharge (ESD)
IEC
61000-4-2
± 6
kV contact
± 8
kV air
± 6 kV
contact
± 8
kV air
Floors
should be wood, concrete or ceramic
tile. If floors are covered with synthetic
mater
ial, the relative humidity should be at
least 30%.
Electrical Fast
Tr
ansient/Burst
IEC
61000-4-4
± 2 kV for power
supply lines
± 1
kV for
input/output lines
±
2 kV for power
supply lines
± 1
kV for input/output
Lines
Mains power quality should be that of a
typical commercial or hospital environment.
Surge
IEC
61000-4-5
±
1 kV line(s) to
line(s)
±
2 kV line(s) to
earth
± 1 kV line(s) to
line(s)
± 2
kV line(s) to earth
Mains power quality should be that of a
typical commercial or hospital environment.
Voltage Dips, Short
Interruption
and
Voltage
Variations
on Power Supply
Input Lines
IEC
61000-4-11
<5%
UT
(>95%
dip in UT)
for 0.5 cycle
40%
UT
(60%
dip in UT)
for 5 cycles
70%
UT
(30%
dip in UT)
for 25 cycles
<5%
UT
(>95%
dip in UT)
for 250 cycles
<5%
UT
(>95%
dip in UT)
for 0.5 cycle
40%
UT
(60
% dip in UT)
for 5 cycles
70%
UT
(30%
dip in UT)
for 25 cycles
<5%
UT
(>95%
dip in UT)
for 250 cycles
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the 2401LM Touchmonitor
for Healthcare Applications requires
continued operation during power mains
interruptions, It is recommended that the
2401LM Touchmonitor for Healthcare
Applications
be powered from an
uninterruptible power supply or a battery.
Power Frequency
(50/60
Hz)
Magnetic
Field
IEC 61000-4-8
3
A/m
3 A/m
Power frequency magnetic fields should be
at levels char
acteristic of a typical location in
a typical commercial or
hospital
environment.
NOTE UT is the A.C. mains voltage prior to application of the test level.